Digital Inputs (optional)

The Protocol Plus control can be run by external inputs wired to the control from an
external source such as a PLC or control panel switches. The external run operation
can Run, Hold or Stop profiles 1 through 7 (profile 8 can not be operated externally).
Refer to the table below for the inputs required for the desired operation. NOTE: A
profile must be created in the program page before trying to run a profile number.

Input 1

Input 2

Input 3

Profile Selected

ON

OFF

OFF

1

OFF

ON

OFF

2

ON

ON

OFF

3

OFF

OFF

ON

4

ON

OFF

ON

5

OFF

ON

ON

6

ON

ON

ON

7

OFF

OFF

OFF

none

To start the selected profile, set Input 1, 2, 3 or 4 to ON.

To hold a profile, set Input 4 to OFF.

To stop a profile, set all inputs to OFF.

Digital Inputs