3
Temperature
Model
LND 1-42
LND 2-11
LND 2-24
Time to Temperature
25
°
C - 100
°
C
(approximate minutes with
25
°
C - 200
°
C
no load)
25
°
C - 343
°
C
6
25
56
5
14
32
3
12
30
Recovery Time Door Open 1 Min.
100
°
C
(approximate minutes with
200
°
C
no load)
343
°
C
<1
3
10
<1
2
7
<1
2
4
Temperature Uniformity at
100
°
C*
200
°
C*
343
°
C*
±1
°
C
±2
°
C
±3
°
C
±1
°
C
±2
°
C
±3
°
C
±1
°
C
±2
°
C
±3
°
C
Minimum Operating Temperature**
with water cooling
(Approximate w/20 C ambient)
without water cooling
55
°
C
97
°
C
55
°
C
100
°
C
60
°
C
110
°
C
Control Stability
±0.3
°
C
±0.3
°
C
±0.3
°
C
Repeatability
±0.5
°
C
±0.5
°
C
±0.5
°
C
Cooldown Time**
343 - 60
°
C
(approximate minutes with no load)
30
35
45
* Figures are based on actual tests in an empty oven. Uniformity can vary slightly depending on unit
and operating conditions.
** 60
°
F Supply water temperature at 120 GPH for LND1-42, 210 GPH for LND2-11 and LND2-24.