4.3. MRC5000 Setup (Optional)
Refer to instructions provided recorder manufacturer for more
specific installation notes.
Temperature is retransmitted to the MRC5000 recorder from the controller. Set up the
recorder by:
1. Ensure hardware jumper JU1 is in place for the 5 VDC setting (Refer to MRC5000
Manual included).
2. Move
Mode
to
PROG/TEST/CAL
to display
Prog
.
3. Press ▼ twice to display
Inps
. Move to each Parameter Code using ▼or ▲. Adjust
each Parameter Code using the settings in Table 2.
4. After adjusting all settings, move
Mode
to
RUN
. Display on both the Recorder and
controller should read the same.
Table 1. MRC 5000 Settings.
Parameter Code
Degrees C
Degrees F
Inps
18
18
Icor
0
0
diSP
On
On
dPOS
0
0
EUU**
400
752
EUL**
0
32
ChUP
400
800*
ChLO
0
0
DFF
1
1
* Change 0-400 chart paper to 0-800 chart paper. Depending on the equipment used, 0-
600 paper may be used if the maximum temperature is 500 degrees F.
** These values must match the settings
RetOutLo
and
RetOutHi
on the Protocol Plus
Control page. For example, if
RetOutLo
is 32,
EUL
must read 32.