4.3. MRC5000 Setup (Optional)


Refer to instructions provided recorder manufacturer for more
specific installation notes.


Temperature is retransmitted to the MRC5000 recorder from the controller. Set up the
recorder by:
1. Ensure hardware jumper JU1 is in place for the 5 VDC setting (Refer to MRC5000

Manual included).

2. Move

Mode

to

PROG/TEST/CAL

to display

Prog

.

3. Press ▼ twice to display

Inps

. Move to each Parameter Code using ▼or ▲. Adjust

each Parameter Code using the settings in Table 2.

4. After adjusting all settings, move

Mode

to

RUN

. Display on both the Recorder and

controller should read the same.


Table 1. MRC 5000 Settings.
Parameter Code

Degrees C

Degrees F

Inps

18

18

Icor

0

0

diSP

On

On

dPOS

0

0

EUU**

400

752

EUL**

0

32

ChUP

400

800*

ChLO

0

0

DFF

1

1

* Change 0-400 chart paper to 0-800 chart paper. Depending on the equipment used, 0-

600 paper may be used if the maximum temperature is 500 degrees F.

** These values must match the settings

RetOutLo

and

RetOutHi

on the Protocol Plus

Control page. For example, if

RetOutLo

is 32,

EUL

must read 32.