Despatch Protocol Plus Controller (Heat/Cool) User Manual

Page 42

background image

36

Test

Menu Item

Display

Default

Range

Setting

Heater output

Test HeatOut

Off

On

High limit relay

Test HiLimOut

Off

On

Relay 1 output

Test Rly1 Out

Off

On

Relay 2 output

Test Rly2 Out

Off

On

Relay 3 output

Test Rly3 Out

Off

On

Relay 4 output

Test Rly4 Out

Off

On

High Limit Sensor

Test HL Temp

(sensor reading)



Zone Cal

Menu Item

Display

Default

Range

Setting

Zone 1 actual

Zone Cal Zone1Act

38°C

-73°C to 760°C (-100°F to
1400°F)

Zone1 displayed

Zone Cal Zone1Dis

38°C

-73°C to 760°C (-100°F to
1400°F)

Zone 2 actual

Zone Cal Zone2Act

260°C

-73°C to 760°C (-100°F to
1400°F)

Zone2 displayed

Zone Cal Zone2Dis

260°C

-73°C to 760°C (-100°F to
1400°F)

Factory calibration*

Zone Cal FactCal

Ready

Ready or Done (push- key)


*only use when no calibration instrument is available


Sensor Cal

Menu Item

Display

Default

Range

Setting

Control Sensor Cal

Ctrl Sens 0F
Ctrl Sens 1000F
Ctrl Sens Done

-100°F
1400°F
Done

-100 to 1400°F
-100 to 1400°F
(read only)

HiLim Sensor Cal

HL Sens 0F
HL Sens 1000F
HL Sens Done

-100°F
1400°F
Done

-100 to 1400°F
-100 to 1400°F
(read only)

Retransmit Cal

RetCalLo XXXX *
RetCalHi XXXX *

0
4096

0 to 4096**
0 to 4096***


* note that the actual RetCalLo and RetCalHi values displayed are of no importance.

** press

keys until retransmission output = 1VDC

*** press

keys until retransmission output = 5VDC