Dakota Ultrasonics

4

The PRB-0 key is used to "zero" the VX in much the same way that a

mechanical micrometer is zeroed. If the gauge is not zeroed correctly, all

of the measurements that the gauge makes may be in error by some fixed

value. Refer to page 12 for an explanation of this important procedure.

The CAL key is used to enter and exit the VX's calibration mode. This

mode is used to adjust the material thickness value that the VX will use

when calculating sound velocity. The gauge will either calculate the sound-

velocity from a sample of the material being measured, or allow a known

velocity value to be entered directly. Refer to page 13 for an explanation of

the two CAL functions available.

The IN/MM key is used to switch back and forth between English and

metric units. This key may be used at any time, whether the gauge is

displaying a thickness (IN or MM) or a velocity value (IN/

µs or M/s).