Dakota Ultrasonics VX User Manual
Page 18
Dakota Ultrasonics
14
Calibration to a known thickness
NOTE: This procedure requires a sample piece of the specific
material to be measured, the exact thickness of which is known,
e.g. from having been measured by some other means.
1) Make sure the VX is on.
2) Perform a Probe-Zero (refer to page 12)
3) Apply couplant to the sample piece.
4) Press the transducer against the sample piece, making sure that
the transducer sits flat against the surface of the sample. The
display should show some (probably incorrect) sound velocity
value, and the Stability Indicator should have nearly all its bars on.
5) Having achieved a stable reading, remove the transducer. If the
displayed value changes from the value shown while the
transducer was coupled, repeat step 4.
7) Press the CAL key. The IN /
µs (or M /s) symbol should begin
flashing.
8) Press the CAL key again. The IN (or MM) symbols should begin
flashing and a thickness value is displayed. Use the UP and
DOWN arrow keys to adjust the displayed thickness up or down,
until it matches the thickness of the sample piece.
9) Press the CAL key once more to exit the calibration mode. The VX
will now display sound velocities of materials of the same thickness
as entered.