Dakota Ultrasonics VX User Manual
Page 19
VX Ultrasonic Velocity Gauge
15
Calibration to a known velocity
NOTE: This procedure requires that the operator know the sound-
velocity of the material to be measured. A table of common
materials and their sound-velocities can be found in Appendix C.
1) Place the transducer on the material to be measured and obtain a
stable reading.
2) Press the CAL key to enter calibration mode. The IN /
µs (or M /s)
symbol will begin flashing.
3) Use the UP and DOWN arrow keys to adjust the displayed velocity
up or down, until it matches the known sound-velocity of the
material.
4) Press the CAL key again and the display will show the thickness
value calculated for the material.
4) Press the CAL key once more to exit the calibration mode. The VX
is now ready to perform measurements.
NOTE: During the calibration procedure when IN /
µs, or M /s is
flashing, pressing the PRB-0 key will restore the gauge to the factory
default sound-velocity for steel (.2330 IN /
µ
s or 5920 M /s).