While the VX excels at making single point measurements, it is

sometimes desirable to examine a larger region to check for material

consistency. The VX includes a feature, called Scan Mode, which allows it

to do just that.

In normal operation, the VX performs and displays four measurements

every second, which is quite adequate for single measurements. In Scan

Mode, however, the gauge performs sixteen measurements every second,

but does not display them. While the transducer is in contact with the

material being measured, the VX is keeping track of the fastest velocity it

finds. The transducer may be "scrubbed" across a surface, and any brief

interruptions in the signal will be ignored. When the transducer loses

contact with the surface for more than a second, the VX will display the

fastest sound velocity it found.

When the VX is not in calibration mode, press the UP arrow key to turn

Scan Mode on and off. A brief message will appear in the display

confirming the operation. While scanning, the display will show a moving

series of dashes instead of a sound value.