Model 6105 Operations Manual Rev I
03/03/2014
Page 9 of 22
Power Requirements
The power cord of this chamber is equipped with a grounded plug. Be sure to connect the ground. CARON
recommends that the chamber have a dedicated wall outlet. Verify correct power supply required for
particular unit. (Nema 6-15 15A Cord, Model 6105-2 only.)
(CEE 7/7 Continental European 16A Cord, 6105-3 only)
Water Connections
Plumb the water drain connection to a facility drain. Connect the water inlet to a water supply. The chamber
requires distilled, reverse osmosis or deionized water with resistivity between 0.05 & 1 Mega Ohmn-cm and
pH above 6.5. Using water outside this range will void the warranty.
Carboy water bottle: BTL101 (optional)
Mount the universal carboy holder either on the top of the chamber or on the side (as shown below). Use the
mounting screw holes provided and secure the holder. Connect tubing to carboy, fill bottle and install as
shown. The water supply must gravity drain. The bottle cap must have a weep hole for it to work properly.
Water outlet from
chamber (drain)
Universal carboy
holder
Water inlet to
chamber (supply)
Carboy
Water outlet from
chamber (drain)
Connect the unit to a dedicated grounded outlet that matches the
units power requirements. Failure to do so could result in
electrical shock or damage to the unit.