MAINTENANCE
The CARON chamber has been robustly designed to minimize performance
problems. However, regular maintenance is very important for continuous trouble
free operation.
As a general rule, CARON recommends an annual calibration check of the
temperature, and humidity systems. CARON offers a full range of on-site calibration
and validation services. We also offer preventative maintenance contracts on our
equipment. Contact our Service department for details at 740-373-6809 or visit us
on the web at
Electrical components (circuit breakers, fuses, relays, power supplies, etc.) are located underneath the
top/back panel. For additional service support, contact your local distributor or CARON service department
at
. When contacting CARON, please provide model number and serial number
Routine Maintenance
Routine chamber maintenance is necessary to keep the chamber working properly. The recommended
chamber maintenance schedule is:
a) Clean interior chamber (as needed)
b) If the humidity control portion of the chamber is going to be unused for 20 days, the small internal
water reservoir should be drained. To drain the internal reservoir:
turn the chamber on
set the humidity switch to off
set the temperature to 100ºC
open the back vent (remove stopper)
run for four hours
leave humidity switch off until humidity control is required
c) Check drains for blockage (monthly)
WARNING: Before removing top/back panel, disconnect all power!