Changing the calibration, high temperature, Cal.h – Carbolite 301 User Manual
Page 17

Customer Calibration
17
MC16-GB-C
– 1.04
8.5
Changing the Calibration, Low Temperature Offset -
OFS.L
OFS.L follows on from CAL.L when the Page key is pressed.
Press the Up or Down arrow key to display the current calibration low offset.
Press the Up or Down arrow key to alter the low offset value.
The value will then be stored without any further input.
8.6
Changing the Calibration, High Temperature
–
CAL.H
CAL.H follows on from OFS.L when the Page key is pressed.
Press the Up or Down arrow key to display the current calibration high temperature.
Press the Up or Down arrow key to change the temperature value.
The value will then be stored without any further input.
8.7
Changing the Calibration, High Temperature Offset -
OFS.H
OFS.H follows on from CAL.H when the Page key is pressed.
Press the Up or Down arrow key to display the current calibration high offset.
Press the Up or Down arrow key to alter the high offset value.
The value will then be stored without any further input.
OFS.L CAL.H
Hold
If starting from Home Display
If password required
CAL.L OFS.L
Hold
If starting from Home Display
If password required
CAL.H OFS.H
Hold
If starting from Home Display
If password required