C.B.S. Scientific SG Series User Manual
Page 23

C.B.S.  Scientific
23
Adjustable Height Sequencer
5.2 Troubleshooting Guide
Problem/Condition
Cause & Solution
ELECTRICAL 
• 'No Load' power supply indicator 
 (no current) 
 
• loose connection at DC source or at terminal: secure if not broken 
• press down firmly on both safety cover lids 
• loose electrode terminal: finger tighten, check for corrosion: clean 
thoroughly 
• break in platinum wire: check for continuos wire 
• if sides of plate sandwich not taped - may be grounding; check for 
leaks/streaking- wipe clean, check number and type of clamps; ensure 
equal number and symmetrical spacing 
• incorrect buffer levels 
• gel hydrolyzed: recheck gel components; 
re-pour 
• 'sparking' near top of gel
• check upper buffer level and refill
• 'sparking' near lower gel chamber
• check lower buffer level: too low or too high
• Over heating of buffer and plates
• buffer: check concentration & date made 
• hydrolyzed gel: date of gel stock solution; remake 
• set power conditions to run gel at approximately 50°C 
• Upper well & gasket area
• not flush & tight against silicone seal: check clamps and bar clamp (if 
used) 
GLASS PLATES 
• cracking at notched area 
 
• check clamps for correct size, Cat.# GPC-0001 
• optional bar clamp: too tight and uneven 
• cracking across main area
• overheating (see ELECTRICAL section above) 
• use aluminum heat dispersion plate (cooling) 
• check clamps: check for proper size; avoid using metal bulldog-type 
clamps 
• check for pre-existing scratches or chips along edges 
LEAKS 
• upper buffer reservoir and sample wells/gasket area 
 
• ensure notched or shorter plate tightly flushed against silicone gasket; 
check for proper plate alignment 
• thoroughly clean silicone gasket area 
• replace gasket if worn, cracked or flattened 
• check tightness & clamp type 
• capillary leakage
• sides of gel plate sandwich (see 'no load' section above) 
• ensure that an equal number of white spring clamps are symmetrically 
placed on both sides of plates 
• check clamp model number (GPC-0001 vs. GPC-0002) 
• ensure stainless steel spring in clamp securely in place 
• space and sequencing comb identical thickness 
• bowed (not flat) plate set 
• Polymerization problem (see following section) 
• gel casting
• ensure spacers meticulously clean 
• sloppy technique if using tape 
• check excess gel solution (if saved) for proper polymerization time 
DEFORMED SAMPLE WELLS 
• deformation when comb removed 
 
• comb inserted too far 
• gel dried out near surface: rinse comb with buffer before removal of 
comb 
• gel not polymerized 
BUBBLES 
 
• plates not cleaned thoroughly prior to casting
SEPARATION OF PLATES AFTER 
ELECTROPHORESIS 
• sticking of gel to one or both plates 
 
 
• plates not cleaned thoroughly 
• plate (1) not silanized (recommend 'GEL COAT' 
• plate pried apart too quickly: separate slowly and carefully; use Wedge 
plate separator, Cat. # WPS-100 
READ FRAMES 
• fuzzy/wavy 
 
• before loading samples, flush urea out of sample wells using pipette 
tip or syringe with bent needle 
• smeary
• re-pour gel using freshly purified gel components
SMILING 
 
• use an aluminum heat dispersion plate 
• avoid loading samples in the two outermost wells on both sides 
• reduce power setting 