C.B.S. Scientific SG Series User Manual
Page 16

C.B.S.  Scientific
16
Adjustable Height Sequencer
5. Verify that the upper buffer reservoir drain is secure. For the single 20cm wide sequencer,
fill the upper buffer reservoir with approximately 350ml of electrophoresis buffer. Make sure 
no buffer is leaking from the upper buffer reservoir. Fill the lower reservoir with 
approximately 350ml of electrophoresis buffer. For the dual 20cm wide sequencer, fill the 
upper reservoir 375ml and the lower with 350ml. For the single 33cm wide sequencer, fill 
the upper buffer reservoir with approximately 550ml of electrophoresis buffer. Make sure no 
buffer is leaking from the upper buffer reservoir. Fill the lower reservoir with approximately 
450ml of electrophoresis buffer. For the dual 33cm wide sequencer, fill the upper reservoir 
450ml and the lower with 450ml. Be sure no bubbles obstruct buffer contact with the lower 
edge of the gel. Note: Do NOT overfill buffer reservoirs or allow buffer to make contact with 
the banana jacks. 
WARNING:
No buffer should leak through or around the silicone gasket or down the side of
the gel assembly. Leakage may allow the upper reservoir to drain dry or cause
arcing and damage to the apparatus.
6. Attach the reservoir safety covers to the upper and lower reservoirs. Connect the female end
of the black power cord into the safety cover of the upper reservoir. Connect the female end 
of the red power cord into the safety cover of the lower reservoir. 
 
7. Be sure to attach Temperature Strip (Cat. # EGT-100) to upper or middle of glass plate, to 
monitor the temperature of the gel.
 
8. Connect the DC power leads to the power supply with the proper polarity. Make sure the 
black leads are connected to the black cathode (-) and the red leads are connected to the red 
anode (+). 
 
9. If using a rectangular comb, rinse sample wells with upper reservoir buffer. Before loading 
samples, pre-electrophorese the gel for 30 to 60 minutes (See Table 5.1 for recommended 
DC power settings). Best results are achieved with a gel surface temperature of 50
°C.
 
 
WARNING:
Excessive power will cause the gel to overheat and crack the glass plates.
 
 
4.2 Loading Samples 
1. Prepare the samples by heating in an appropriate loading buffer to 95
°C to 100°C for 3-5
minutes, vortexing, and then chilling immediately on ice.
 
2. At the end of the pre-electrophoresis period. TURN OFF THE POWER SUPPLY AND 
DISCONNECT BOTH DC POWER CORDS FROM THE POWER SUPPLY. Remove power 
cords from both reservoirs. Open upper reservoir safety cover. 
 
3. Immediately prior to loading the samples, rinse the wells of the gel thoroughly with 
electrophoresis buffer. Use a sequencing loading tip or a syringe with a bent needle to wash 
away urea that has diffused into the wells. Load samples in a loading pattern A C G T. You 
can use a marker to draw on the proper loading lanes. For sharktooth combs load 2-3ul of 
sample. (For rectangular tooth combs see comb volumes, Table 5.3).
 
 
 
 
 
 
 
 
 
 