beautypg.com

C.B.S. Scientific EPS – 200 X User Manual

Page 8

background image

EPS-200 X

6

1. Connect the main power cord: Connect the main power cord to the power cord receptacle

on the rear panel and to a suitable grounded AC power outlet.

2. Connect the HV power leads: Connect the apparatus to the power supply by plugging the

lead connectors into one pair of the recessed output jacks. The power leads are

non-reversible as per to the color coordinated to the output jacks.

3. Turn on the main power switch: Turn on the main power switch on the rear panel. The

front panel and the main power LED indicator should be illuminated.

4. Select constant mode: Press the CONST button to toggle between constant voltage and

constant current mode. The selection is indicated by the green LED.

5. Adjust limit: Press up and down buttons to set the constant variable to the desired value.

6. Check other run variable: Press the MODE button to check and verify the set value of the

other (non-constant) run variable. This variable is set by default to the maximum output of

the power supply. This is the normal condition or setting for simple electrophoretic

separations. Optional: If there is another upper limit desired for this non-constant variable,

use the up and down buttons to set it to the desired value.

7. Set the timer (Optional): Press the MODE button to move the green indicator LED light to

under clock symbol.

Continuous run: Leave the timer at “0” to start a continuous run. The power supply will

continue to operate indefinitely, until stopped manually. The timer will start counting up,

indicating the length of the run in minutes. After reaching the maximum time of 999 minutes

the timer will automatically reset and recount again from 0. Therefore for over night

operation the resulting operation time is the addition of the current timer value with the

previous maximum timer 1440 minutes.

Timed run: Press the up and down buttons to set the timer to the desired value (minutes).

The timer counts down during the run, indicating the time remaining. At the end of the timed

run, a ~3 second beep sounds and the display changes to “OFF”. To reset the power supply

press any button except the RUN button.