10
E7
Short Circuit
Load current exceeded 2 A
1. Check for any short circuit or evidence
of electrical arching.
2. Buffer concentration is too high.
3. Press any button* to clear the error
code.
*Press any button, except the RUN button, to clear the error message.
D-3. Replacing the fuses
1. Unplug the power cord from the power supply.
2. Locate the fuse compartment on the rear panel.
3. Use a screwdriver to pry out the fuse compartment. There are two fuses in the compartment.
Check and replace any blown fuses with a fuse of the same size and ratings.
4. FUSES (2); 250 V, 3.15 A, 5 x 20 mm fast blow.
                    See also other documents in the category C.B.S. Scientific Equipment:                                                            
                    - FCU-1000 (24 pages)
 - EBU-204 (16 pages)
 - EBX-700 (16 pages)
 - EBU-302 (12 pages)
 - EBU-102 (16 pages)
 - EBU-222 (12 pages)
 - EBU-3000 (16 pages)
 - GSL-1001 (7 pages)
 - GM-20 (12 pages)
 - MGU-303 (12 pages)
 - MGU-102T (16 pages)
 - MGU-102T-FL (16 pages)
 - MGU-402T (16 pages)
 - MGU-402T-FL (16 pages)
 - WMGE-600 (20 pages)
 - SGU-014T-02 (16 pages)
 - GCSGU-014T-02 (16 pages)
 - SGE-014-02 (16 pages)
 - GCSGE-014-02 (16 pages)
 - ACGT-16520 (4 pages)
 - ECU-040-20 (16 pages)
 - HTLE-7002 (44 pages)
 - M-312 (12 pages)
 - LPE-4003 (16 pages)
 - DTSK-2001-110 (60 pages)
 - DGGEK-1001 (36 pages)
 - DGGEK-4001 (19 pages)
 - TTGEK-2001-110 (60 pages)
 - P-030-02 (21 pages)
 - EPS – 300 X (15 pages)
 - TTGE-2001 (16 pages)
 - EPS - 600 (15 pages)
 - EPS-3000P-II (14 pages)
 - DDH Series (40 pages)
 - DASG Series (36 pages)
 - SG Series (20 pages)
 - DCX-700 (24 pages)
 - QNC-700 (24 pages)
 - DCX-800C (24 pages)
 - DCX4K Series (3 pages)
 - MGV-102 (24 pages)
 - MGV-102-20 (20 pages)
 - ASG-250 (24 pages)
 - DSG Series (20 pages)
 