SECTION 1
General Information
1.1 Introduction
Flip Lid Horizontal Mini-Gel systems MGU-102T-FL and MGU-202T-FL provide a
compact format to rapidly separate and identify DNA and RNA fragments, PCR
products and synthetic oligonucleotides, while using minimal quantities of expensive
reagents. The tapered baffle system allows gel casting directly onto the unit or remov-
able trays without having to seal with agarose plugs or tape. The gel bed and gel tray
are UV transparent. Combs are offered in several thicknesses and formats to accom-
modate a wide variety of experimental applications.
1.2
Specifications
Constructions
Buffer chamber, lid
Acrylic
Electrodes
Platinum wire .010” diameter
Power cords
FR Silicone rated 7500VDC, 200mA, 65ºC
Tapered baffles
Polycarbonate
Combs
Polycarbonate, Anodized Aluminum or Teflon
Comb backing
Acrylic
UVT Gel Tray (optional)
U-V transparent acrylic
Safety Certification
EN61010-1-1993 (IEC1010-1)
*Calculated using gel thickness of 0.5cm
1.3
Safety
Power to the Mini-Horizontal units is to be supplied by an external DC voltage power
supply that must be ground isolated so that the DC voltage output floats with respect
to ground. For any power supply used, the maximum specified operating parameters
for the units are:
Maximum Limits
250VDC voltage
15 watts power
60 mA current
50*C ambient temperature
Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. Current to the unit
is broken when the lid is removed. Do not attempt to use the unit without the safety
lid. Always turn the power supply off before removing the lid, or when working with
the unit in any way. Follow safety precautions specified by the power supply manu-
facturer.
Model #
MGU-102T-FL
MGU-202T-FL
Shipping weight
3 lb.
3 lb.
Overall Size W x L x H (cm)
14 x 19.7 x 12.7
17 x 19.7 x 12.7
Gel Bed Size W x L (cm)
5.5 x 8.5
7.5 x 10
Recommended Buffer Volume
150ml
250ml
Agarose Solution Required*
23ml
36ml
# of Slots in Gel Tray
2
2
Pathlength/Slot (cm)
5.0
5.5
Voltage limit
250VDC
250VDC