Ii. multi-channel model, Research – part b – maintenance – Eppendorf Research Family - fix, variable, multi User Manual

Page 26

background image

86

A Replacing the O-rings

– Hold down the ejection button on the side of the pipette and unscrew the

lower part counterclockwise from the grip.

– Using your index finger, press the tool from behind against the nose cone.

Using your thumb, apply pressure from the other side.

– The O-ring is severed at one point and can be removed from the nose cone.

Attaching a new O-ring

– Attach the assembly aid (shortened pipette tip) onto the nose cone.

– Slide the new O-ring over the tip and onto the nose cone.

B Replacing the piston seals

– Hold down the ejection key and unscrew the lower part (Fig. 2-6)

counterclockwise from the grip.

– With the cone facing downwards, place the pipette lower part horizontally

onto a table and press it down slightly.

– Press side B of the tool into the opening (Fig. 2-8) in the housing (Fig. 2-9).

The two stop hooks on the inside are depressed and the cover
plate (Fig. 2-7) pops up!

– Push opening A of the tool

(suitable for the pipette size at
hand) from below over the
nose cones in such a way as
to ensure that the sharp edge
in the opening of the tool is
overlapping with the O-ring
(see Fig. 6).

II. Multi-channel model

C

A

B

Fig. 6

Research – Part B – Maintenance