2 innova 2350, Innova 2350 – Eppendorf 2350 Biological Shakers Innova User Manual
Page 28

Technical data
New Brunswick
™
Innova
®
2300/2350 Shakers
English (EN)
28
7.1.2
Innova 2350
Tab. 7-2:
Shaking
Contruction
• Heavy gauge steel
• Phosphate coated
• Textured painted frame
Weight
Net
• 49
kg
(108
lb)
Gross
• 66
kg
(145
lb)
Dimensions
Width
74
cm
(29
in)
Depth
56.5
cm (22 1/4
in)
Height
17
cm (6 3/8
in)
Platform dimensions
Width
46
cm
(18
in)
Depth
76
cm
(30
in)
Speed
• 25
rpm
–
400
rpm
• 25
rpm
–
300
rpm
Motion
• 25.4
mm
(1
in)
diameter
circular
orbit
• 50.8
mm
(2
in) diameter circular
orbit
Indication
• LED digital electric display
• 1
rpm
increments
Setpoint and
control
• Digital adjustment with PI microprocessor control and instantaneous visual
feedback
Accuracy
• ±1
rpm
At 25
–
400
rpm, the unit will perform to specifications with up to ±10
% line voltage
fluctuation. To attain speed accuracy at 401
–
500, the line voltage cannot be lower than 5 %
of the rated voltage.
Drive
• Eccentric counterbalanced ball bearing drive
Keypad timer
• Programmable shaking periods from 0
hr
1
min
–
99.9
hr by a digital timer
• Timer shuts off at the end of shaking period and energizes status light
• Timer counts down and digital display indicates remaining time
• Can be deactivated for continuous operation
• Unit will display total accumulated running time for service information
Operating ambient
environment
• 0
°C
–
60
°C
• 90
%
• Non-condensing
• Up to 2000 m