Technical data
New Brunswick
™
Innova
®
2300/2350 Shakers
English (EN)
27
7
Technical data
7.1
Specifications
7.1.1
Innova 2300
Tab. 7-1:
Shaking
Speed
• 25
rpm
–
500
rpm
• 25
rpm
–
300
rpm
Motion
• 25.4
mm
(1
in)
diameter
circular
orbit
• 50.8
mm
(2
in) diameter circular
orbit
Indication
• LED digital electric display
• 1
rpm
increments
Setpoint and
control
• Digital adjustment with PI microprocessor control and instantaneous visual
feedback
Accuracy
• ±1
rpm
At 25
–
400
rpm, the unit will perform to specifications with up to ±10
% line voltage
fluctuation. To attain speed accuracy at 401
–
500, the line voltage cannot be lower than 5 %
of the rated voltage.
Drive
• Eccentric counterbalanced ball bearing drive
Keypad timer
• Programmable shaking periods from 0
hr
1
min
–
99.9
hr by a digital timer
• Timer shuts off at the end of shaking period and energizes status light
• Timer counts down and digital display indicates remaining time
• Can be deactivated for continuous operation
• Unit will display total accumulated running time for service information
Operating ambient
environment
• 0
°C
–
60
°C
• Up to 90
%,
non-condensing
• Up to 2000 m
Self-diagnostic
status
• Warning signal (audible and visible) indicates when shaking speed deviates
more than 5 rpm
• Warning signal (audible and visible) indicates when timer operation has expired
• Audible alarm can be deactivated/reactivated by the operator
Remote speed
monitoring
(optional)
• Chart recorder output for speed 0.5
V, 1 V per 100 rpm
• Accuracy ±25
mV
Automatic restart
• Unit will automatically restart after undesired power interruption
• Setpoints are maintained by non-volatile memory
• Interruption is indicated by a flashing LED
Motor
• 1/8
HP
• 3 phase
• Brushless ball bearing DC motor
Electrical service
• 100
V,
120
V,
220
V,
240
V
• All voltages 50/60
Hz
150
VA
• Universal power entry system adapts to U.S. or international requirements
Electrical
protection
• Main fuse(s) in power entry module
• Control circuits provided with separate fuse