3 universal power module, 4 control panel, Universal power module – Eppendorf 2350 Biological Shakers Innova User Manual
Page 12: Control panel

Product description
New Brunswick
™
Innova
®
2300/2350 Shakers
English (EN)
12
3.3
Universal power module
The power module contains a voltage card and fuse holder which are used to select the appropriate voltage
and fusing. This universal power entry system adapts to worldwide power requirements. Voltage and fusing
has been set prior to shipment. Innova shakers are available in 100
V,
120
V,
220
V and 240
V
configurations that accommodate both 50 and 60
Hz frequencies.
3.4
Control panel
The control panel is located on the front of the instrument. It serves as the operator interface, and consists
of:
• A START/STOP key
• An up arrow (
)
• A down arrow (
)
• A SELECT key
• A 3-digit LED display that provides numeric values, and some letter codes
• 4 function indicator lights
• 4 status indicator lights
WARNING! Risk of electric shock and/or damage to device!
Do not plug the shaker into a power source until you check the voltage and frequency
settings.
Check the voltage selection on the power entry module against the main identification
label or serial tag, located on the side or rear of the unit.