5 maintenance, 1 cleaning, Eppendorf smartblock — instructions for use – Eppendorf SmartBlock User Manual
Page 7
7
Eppendorf SmartBlock — Instructions for use
SmartBlock plates, SmartBlock PCR 96 or SmartBlock PCR 384:
5
Maintenance
5 Maintenance
5.1 Cleaning
Auxiliary equipment
•
Lint-free cloth
•
Mild, soap-based lab cleaner
•
Dist. water
Clean the thermoblock immediately if sample fluid enters the bore holes or comes into
contact with the surfaces.
1. Clean with a mild soap solution.
2. Wipe off the soap solution with dist. water.
3. Dry the cleaned thermoblock.
To ensure uniform temperature control,
place the lid on the thermoblock.
1 height sensor
1
NOTICE!
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories,
such as strong and weak bases, strong acids, acetone, formaldehyde,
halogenated hydrocarbons or phenol.
If the device becomes contaminated with aggressive chemicals, clean it
immediately with a mild cleaning agent.
NOTICE!
NOTICE! Corrosion from aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive
polishes.