3 safety, 1 intended use, 2 warning signs on the device – Eppendorf SmartBlock User Manual
Page 4: 4 installation and operation, 1 installing the thermoblock, Eppendorf smartblock — instructions for use
![background image](/manuals/551354/4/background.png)
4
Eppendorf SmartBlock — Instructions for use
3
Safety
3 Safety
3.1 Intended use
Thermoblocks for the ThermoMixer C and ThermoStat C (SmartBlocks) devices are
intended for use in a molecular biology laboratory.
The thermoblocks are used to hold tubes and plates during sample preparation and
during temperature control and mixing of samples.
The thermoblocks are intended exclusively for indoor use. All country-specific safety
requirements for operating electrical equipment in the laboratory must be observed.
Only use Eppendorf accessories or accessories recommended by Eppendorf.
The thermoblocks must only be operated by appropriately qualified and instructed
personnel.
3.2 Warning signs on the device
4
Installation and operation
4 Installation and operation
4.1 Installing the thermoblock
Representation Meaning
Location
Risk of burns from hot surfaces.
On the ther-
moblock
Caution! Risk of injury from moving parts.
Observe the operating manual.
On the ther-
moblock
WARNING!
WARNING! Personal injury or material damage due to chemically or
mechanically damaged thermoblocks.
Do not use thermoblocks that show signs of corrosion or mechanical
damage.
Regularly check the condition of the thermoblocks.
WARNING!
WARNING! Contamination due to opening seals of consumables.
In the following cases, the seals of tubes can spring open. Sample material
can escape.
•
high vapor pressure of the content
•
improperly sealed cover
•
damaged sealing lip
•
improperly fastened foil
Always check that consumables have been sealed tightly before use.