beautypg.com

6 quick start, 7 empty battery, Quick start – Eppendorf Repeater plus User Manual

Page 10: Empty battery

background image

Multipette® plus · Repeater® plus — Operating manual

10

4.6 Quick Start

1.

Select Combitip.

2.

With Combitip 25, 50 mL: connect adapter to the Combitip.

3.

Insert Combitip (see p. 8).

4.

Fill Combitip:
a

Immerse the tip of the Combitip in the liquid.

b

Slowly draw up the filling lever.

5.

Reverse stroke:
a

Press the dispensing lever.

b

Discard the first dispensing step (reverse stroke).

6.

Dispense:
a

Place the tip of the Combitip against the wall of the tube.

b

Press the dispensing lever.

7.

Discard the Combitip:
a

Push the filling lever right to the bottom.

b

Press both ejection keys at the same time.

4.7 Empty Battery

The Multipette plus/Repeater plus can still function if the electronics

fail or the battery is empty.
To set volumes, use the volume table (see p. 7) or proceed as follows

in order to determine the position required on the volume selection

wheel:
1. Read the volume for position 1 on the volume wheel from the

Combitip (fig. 2-1).

2. Divide the desired volume by the volume with position 1.
Examples for Combitip

Hint!

Only trained staff already familiar with the operating man-

ual and the Multipette plus/the Repeater plus may work to

these short instructions. Follow the safety instructions.

Volume for position 1
Desired volume of dosage
Calculation
Position to be set

100 μL
200 μL
200 / 100 = 2
2

Desired volume of dosage
Calculation
Position to be set

750 μL
750 / 100 = 7.5
between 7 and 8

Desired volume of dosage
Calculation
Position to be set

50 μL
50 / 100 = 0.5
before 1

This manual is related to the following products: