7Technical data
7
Technical data
The data is valid for the following test conditions, in accordance with
ISO 8655:
8Ordering Information
8
Ordering Information
Combitip
advanced
Testing
volume
Error limits
Error
Systematic
error
Random
error
± %
± μL
± %
± μL
0.1 mL
white
2 μL
±1.6 ±0.032
±3.0
±0.06
20 μL
±1.0 ±0.2 ±2.0
±0.4
0.2 mL
light blue
4 μL
±1.3 ±0.052
±2.0
±0.08
40 μL
±0.8 ±0.32 ±1.5
±0.6
0.5 mL
violet
10 μL
±0.9 ±0.09 ±1.5
±0.15
100 μL
±0.8 ±0.8 ±0.6
±0.6
1 mL
yellow
20 μL
±0.9 ±0.18 ±0.9
±0.18
200 μL
±0.6 ±1.2 ±0.4
±0.8
2.5 mL
green
50 μL
±0.8 ±0.4 ±0.8
±0.4
500 μL
±0.5 ±2.5 ±0.3
±1.5
5 mL
blue
100 μL
±0.6 ±0.6 ±0.6
±0.6
1 000 μL
±0.5 ±5.0
±0.25
±2.5
10 mL
orange
200 μL
±0.5 ±1.0
±0.6
±1.2
2 000 μL
±0.5 ±10
±0.25
±5.0
25 mL
red
500 μL
±0.4 ±2.0
±0.6
±3.0
5 000 μL
±0.3 ±15
±0.25
±12.5
50 mL
light grey
1 000 μL
±0.3 ±3.0
±0.5
±5.0
10 000 μL
±0.3 ±30
±0.3
±30
Liquid
Water, double distilled, degased
Test temperature
20 to 25 °C, constant to ±0.5 °C
with original Eppendorf Combitips
Number of determinations 10, before dispensing, Combitip com-
pletely filled
Operating temperature
+4 °C to +40 °C
Battery
Lithium button cell
3 V
210 mAh
Service life about 2 years
CAUTION!
CAUTION! Poor safety due to incorrect accessories
and spare parts.
The use of accessories and spare parts other than those
recommended by Eppendorf may impair the safety, func-
tion and precision of the device. Eppendorf cannot be held
liable or accept any liability for damage resulting from the
use of incorrect or non-recommended accessories and
spare parts or from the improper use of such equipment.
Only use accessories and original spare parts