System management

Mastercycler pro

English (EN)

59

Performing a
software
synchronization

Use the

update

node to perform a software synchronization between the cycler and the

control panel. The software synchronization may be necessary if you receive error
messages after updating the software of the control panel.

1. Highlight

Update

node.

2. Press

Cycler

function key.

3. Select the Cycler.
4. Press

Start

.

Exporting
user-related data

Use the

UserLog

node to export or display all the actions that were performed on the

control panel as a log file.
You can save login and logout information, all changes and interventions in a running
program related to specific users.

 Mark

UserLog

and press the desired function key:

Show

: Display user log. You can print or export the user log in the display.

Exp+Clear

: Transfer the user log to the MMC or USB storage medium and delete

the log from the control panel afterwards.

Export

: Copy the user log to the MMC or USB storage medium.