4 information on product liability, Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 42

Centrifuge 5804/5804 R/5810/5810 R — Operating manual
42
For all other rotors and rotor lids of this centrifuge there is no service life limit as long as the
following requirements are met:
•
Correct use,
•
Recommended maintenance
•
Undamaged condition
The date of manufacture is stamped on the rotors in the format 03/12 (= March 2012) or on the
inside of the plastic rotor lids in the form of a clock .
To ensure aerosol tightness, the following applies:
•
Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.
•
Replace the seal of QuickLock rotor lids after 50 autoclaving cycles.
3.4
Information on product liability
In the following cases, the designated protection of the device may be compromised. Liability for
any resulting property damage or personal injury is then transferred to the operator:
•
The device is not used in accordance with the operating manual.
•
The device is used outside of its intended use.
•
The device is used with accessories or consumables which are not recommended by
Eppendorf.
•
The device is maintained or repaired by people not authorized by Eppendorf.
•
The user makes unauthorized changes to the device.
Rotor
Maximum service from commissioning onward
A-2-DWP-AT
100,000 mechanical cycles
7 years
A-2-DWP
34,000 mechanical cycles
7 years
A-4-44
34,000 mechanical cycles
7 years
A-4-62
40,000 mechanical cycles
7 years
A-4-81
100,000 mechanical cycles
7 years
F-34-6-38
75,000 mechanical cycles
7 years
FA-45-6-30
7 years
FA-45-48-11
75,000 mechanical cycles
7 years
FA-45-20-17
75,000 mechanical cycles
7 years
F-35-48-17
75,000 mechanical cycles
7 years
S-4-72
60,000 mechanical cycles
7 years
S-4-104
100,000 mechanical cycles
7 years
T-60-11
7 years
Accessories
Maximum service from commissioning onward
Aerosol-tight rotor lid, without replaceable
seals
50 autoclaving cycles
–
Rotor lid QuickLock
3 years
Seals of the QuickLock rotor lid
50 autoclaving cycles
–
Rotor lid and caps made of polycarbonate
(PC), polypropylene (PP) or polyetherimide
(PEI)
50 autoclaving cycles
3 years
Adapters
–
1 year
EN
Operating manual