4 rotors, 1 rotor a-4-81 (only 5810/5810 r), Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 13
![background image](/manuals/551247/13/background.png)
13
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
2.4
Rotors
You can operate the 5804/5804 R/5810/5810 R centrifuge with the following rotors. Before use of
sample tubes, please note the manufacturer's specifications with regard to centrifugation
resistance (max. rcf).
2.4.1
Rotor A-4-81 (only 5810/5810 R)
Rotor A-4-81 with 500 mL rectangular bucket
Hint!
Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.
Only use rotors which bear the name of the centrifuge (e.g. 5804 R).
Max. g-force:
3,220 × g
Rotor A-4-81
Rectangular bucket
500 mL
Aerosol-tight cap
Max. speed:
4,000 rpm
Swing-bucket rotor
with 4 × 500 mL
rectangular buckets
Max. load per bucket
(adapter, tube and
contents):
780 g
Vessel
Vessel
Adapters
Adapter bottom
shape
Max. g-force
Capacity
Order no.
(international)
Tube diameter
Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Vessel
5810 745.004
flat
2,950 × g
1.5/2 mL
Ø 11 mm
4,000 rpm
20/80
43 mm/43 mm
16.5 cm
Blood collection tube
5810 746.000
flat
3,000 × g
1.2 to 5 mL
Ø 11 mm
4,000 rpm
20/80
108 mm/108 mm
16.8 cm
Vessel
5810 720.001
flat
3,000 × g
2.6 to 5 mL
Ø 13 mm
4,000 rpm
25/100
107 mm/108 mm
16.8 cm
Vessel
5810 747.007
flat
3,000 × g
2.6 to 7 mL
Ø 13 mm
4,000 rpm
18/72
108 mm/108 mm
16.8 cm
EN
Operating manual