Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 30
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
30
Blood collection tube
5804 739.000
Round
15,442 × g
4 mL to 7 mL Ø
13 mm
18,353 × g
3/18
107 mm
11.4 cm
Vessel
5804 771.001
Round
15,150 × g
7 to 15 mL
Ø 16 mm
18,000 × g
2/12
112 mm
11.2 cm
Conical tube
5804 776.003
conical
14,450 × g
15 mL
Ø 17.5 mm
17,200 × g
1/6
123 mm
10.7 cm
Vessel
5804 772.008
Round
14,750 × g
15 to 18 mL
Ø 18 mm
17,550 × g
1/6
123 mm
10.9 cm
Vessel
5804 773.004
Round
14,900 × g
20 mL to 30 mL Ø
26 mm
17,700 × g
1/6
123 mm
11.0 cm
Vessel
5804 774.000
Round
15,157 × g
50 mL
Ø 29 mm
18,014 × g
1/6
123 mm
11.2 cm
Conical tube
5804 775.007
conical
14,600 × g
50 mL
Ø 29.5 mm
17,400 × g
1/6
121 mm
10.8 cm
Vessel
-
-
15,550 × g
85 mL
Ø 38 mm
18,500 × g
-/6
121 mm
11.5 cm
Vessel
Vessel
Adapters
Adapter bottom
shape
Max. g-force at 11,000 rpm
(5804/5804 R/5810)
Capacity
Order no.
(international)
Tube diameter
Max. g-force at 12,100 rpm
(5810 R)
Vessels per
adapter/rotor
Max. tube length
with rotor lid
Centrifugation radius
EN
Operating manual