Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 73
73
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
Deceleration times according to DIN 58 970
Tab.
2:
Approximate deceleration times of the different rotors for the levels 0 to 9 (in seconds)
for 230 V devices
These values are to be considered as guidelines. Level 9 means "strongest braking", level 0
means "free deceleration". Considerable fluctuations can occur depending upon the condition of
the device and the load. The deceleration times for the 230 and 120 V devices are almost
identical.
5804/
5804 R
5810/5810
R
Rotor
0
1
2
3
4
5
6
7
8
9
-
A-4-81
532
189
174
143
131
109
95
85
59
31
-
A-4-81- MTP/
Flex
643
191
174
142
131
110
94
83
58
30
-
A-4-62
740
190
170
140
130
110
95
85
55
26
-
A-4-62-MTP
620
190
170
140
130
110
95
85
55
26
A-4-44
470
300
270
220
200
140
100
75
45
23
-
A-2-DWP-AT
857
231
202
176
157
135
μL
112
102
69
39
A-2-DWP
304
174
130
118
100
75
51
44
32
14
FA-45-6-30
759
423
322
231
205
178
148
113
93
58
F-34-6-38
880
370
280
190
170
150
125
95
75
54
FA-45-30-11
240
140
70
45
35
30
25
22
19
18
F-45-30-11
240
140
70
45
35
30
25
22
19
18
F-45-48-PCR
169
119
60
41
31
26
22
19
17
16
T-60-11
800
280
140
95
70
55
45
40
36
36
-
S-4-104
680
192
169
147
130
112
68
46
37
32
S-4-72
360
238
191
168
148
117
56
32
21
15
FA-45-48-11
454
269
141
98
76
62
46
40
35
31
FA-45-20-17
550
342
178
123
97
80
58
51
45
40
F-35-48-17
16
26
40
72
140
185
211
243
304
228
EN
Operating manual