beautypg.com

Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 68

background image

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

68

No rotor

Centrifuge does not start
up.

Error in the drive or in the rotor detection.

Switch the centrifuge off and back on
again after > 20 s.

Press Open

Centrifuge lid could not be locked.

1. Press the open key.
2. Try again to close centrifuge lid.

Close lid

Centrifuge lid not closed properly.

Close the centrifuge lid firmly.

Lift lid

The centrifuge lid does
not open.

The centrifuge lid cannot open automatically.

Lift the centrifuge lid manually.

IMBAL

The centrifuge shakes
when it starts up and
switches off.

Rotor asymmetrically loaded.

Load the rotor symmetrically (see p. 51).

ROTOR

The centrifuge shakes
when it starts up and
switches off.

Rotor not screwed sufficiently tight.

1. Tighten the rotor nut (see p. 51).
2. Check the rotor cone and motor shaft for

grooves and damage.

ROTOR

The centrifuge shakes
when it starts up and
switches off.

Centrifuge was pushed.

Table is not stable.

Position the centrifuge on a stable table
(see p. 48).

SPEED

Centrifuge switches off.

Nominal speed for rotor too high.

Enter the appropriate nominal speed
(see p. 13).

change rotor

The maximum service life of the rotor has
been reached. The warning is displayed after
98,000, 99,000 and 99,600 runs (3 times
after each run). After 100,000 runs, it is
displayed after every run.

Contact Technical Service.

Temperature display
flashes.
(only 5804 R/5810 R)

Temperature deviation from the set value: ±3
°C.

Check the settings.

Wait until the target temperature has been
reached.

Check unhindered air circulation through
the air slots.

Thaw ice or switch off device and allow it
to cool down.

Overtemp

(only 5804 R/5810 R)
Centrifuge switches off
and issues a warning
tone.

Temperature deviation from set value in the
rotor chamber: ±5 °C.

Check the settings.

Check unhindered air circulation through
the air slots.

Thaw ice or switch off device and allow it
to cool down.

Clear memory

Program memory full.

Delete some programs (see p. 61).

Interrupt

Power failure during a run.

Check the mains connection.

Error 1

Error in speed measuring system.

If this error message appears again, test
with a different rotor.

Error 2

Imbalance sensor faulty.

Repeat the run.

Symptom/message

Cause

Remedy

EN

Operating manual

This manual is related to the following products: