9 rotor f-45-48-pcr, 10 rotor t-60-11, Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 32
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
32
2.4.9
Rotor F-45-48-PCR
2.4.10 Rotor T-60-11
Rotor F-45-48-PCR
Fixed-angle rotor for tube strips or
0.2 mL PCR tubes
Max. g-force:
15,294 × g
Max. speed:
12,000 rpm
Max. load (tube and
contents):
6 × 3.5 g
Vessel
Vessel
Adapters
Max. g-force
Capacity
Tube diameter
Max. speed
Vessels per
adapter/rotor
Centrifugation radius
8-tube/5-tube tube
strips
-
15,294 × g
8/5 × 0,2 mL
Ø 6 mm
12,000 rpm
-/6 × 8 and/or -/6 × 5
9.5 cm
Vessel
-
15,294 × g
0.2 mL
Ø 6 mm
12,000 rpm
-/48
9.5 cm
Rotor T-60-11
Drum rotor for tubes
Max. g-force:
14,000 × g
Max. speed:
16,435 rpm
Max. load (tube and
contents):
6 × 70 g
Vessel
Vessel
Adapters
Max. g-force
Capacity
Tube diameter
Max. speed
Vessels per
adapter/rotor
Centrifugation radius
Vessel
-
16,435 × g
1.5/2 mL
Ø 11 mm
14,000 rpm
10/60
7.5 cm
Vessel
-
16,435 × g
0.4 mL
Ø 6 mm
14,000 rpm
20/120
7.5 cm
EN
Operating manual