11 rotor s-4-104 (only 5810/5810 r), Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 33
33
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
2.4.11 Rotor S-4-104 (only 5810/5810 R)
Max. g-force:
3,214 × g
Rotor S-4-104
Round bucket
750 mL
Aerosol-tight cap
Max. speed:
3,900 rpm
Swing-bucket rotor
with 4 × 750 mL
round buckets
Max. load per bucket
(adapter, tube and
contents):
1,000 g
Vessel
Vessel
Adapters
Adapter bottom
shape
Max. g-force
Capacity
Order no.
(international)
Tube diameter
Max. speed
Vessels per
adapter/rotor
Max. tube length
with/without
aerosol-tight
bucket cap
Centrifugation radius
Vessel
5825 740.009
continuous
3,197 × g
1.5/2 mL
Ø 11 mm
3,900 rpm
62/248
39 mm
18.8 cm
Vessel
5825 739.000
continuous
3,214 × g
5 mL
Ø 17 mm 3,900 rpm
14/56
60 mm
18.9 cm
Vessel
5825 738.004
round
3,044 × g
4 to 8 mL
Ø 13 mm × 100 mm
3,900 rpm
23/92
108 mm/115 mm
17.9 cm
Vessel
5825 736.001
round
3,061 × g
7,5 to 12 mL
Ø 16 mm × 98 mm
3,900 rpm
20/80
114 mm/119 mm
18 cm
Vessel
5825 736.001
round
3,061 × g
8 to 16 mL
Ø 16 mm
3,900 rpm
7/28
Load inner bores
only (Fig. 5 on p. 54).
(Do not use the
aerosol-tight cap.)/
125 mm
18 cm
Vessel
5825 743.008
round
3,044 × g
9 mL
Ø 17.5 mm ×
100 mm
3,900 rpm
20/80
106 mm/111 mm
17.9 cm
Conical tube
5825 734.009
conical
3,197 × g
15 mL
Ø 17 mm × 104 mm
3,900 rpm
14/56
120 mm/125 mm
18.8 cm
EN