Maintenance

Eppendorf Eporator®

English (EN)

27

7

Maintenance

7.1

Cleaning

Wet a cloth with a mild cleaning fluid and demineralized water and remove the
contamination on the outside of the device.

7.2

Replacing fuses

The fuse holder is located between the mains/power cord socket and the mains/power
switch.

1. Unplug the mains/power plug.

2. Press the upper and lower end of the plastic springs 1 together and pull the fuse holder

2 fully out.

3. Replace faulty fuses and reinsert the fuse holder. Make sure that the guiding rail 3 is

positioned correctly.

DANGER! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect it from the power supply before any
maintenance or cleaning work is carried out.

Do not allow any liquids to enter the inside of the housing.

Do you use any spray disinfectants on the housing.

Only reconnect the device to the power supply once it is completely dry.

NOTICE! Damage from using aggressive chemicals.

Do not use any aggressive chemicals on the device or its accessories, such as
strong and weak bases, strong acids, acetone, formaldehyde, halogenated
hydrocarbons or phenol.

If the device becomes contaminated with aggressive chemicals, clean it
immediately with a mild cleaning agent.