beautypg.com

5 emptying the emission condenser, 5 usage notes for rotors, 1 rotor a-2-vc – Eppendorf Concentrator plus User Manual

Page 27: Emptying the emission condenser, Usage notes for rotors 5.5.1, Rotor a-2-vc, Ired (see emptying the emission condenser on, 5 operation, Operat ing manual

background image

29

5 Operation

Concentrator plus / Vacufuge® plus — Operating manual

5.4.5

Emptying the emission condenser

1. Depending on the design, there are two options:

The emission condenser is mounted at the side of the device: detach the tube from
the connection and empty the liquid via the upper connection into a collecting vessel.

The emission condenser is mounted at the front of the device: here you can
alternatively remove the black plug in the vessel using a screwdriver and fit a special tap.
Via this tap, you can then empty the liquid directly into a collecting vessel.

2. Dispose of the liquid in accordance with the current legal requirements and regulations for

your application.

5.5

Usage notes for rotors

5.5.1

Rotor A-2-VC

Transferring rotor

CAUTION!

Caution! If handled incorrectly, the rotor can fall over.
The rotor buckets A-2-VC cannot be used as handles.

Before moving the rotor, remove the buckets.

Always pick up the rotor at the rotor cross, using both hands.

CAUTION!

Caution! Wait for rotor to stop.
When the plates are fully loaded, it can happen during the overrun phase of the A-2-VC rotor that
the lid of the device can be opened before the rotor stands still.

Always wait until the rotor stands still before opening the lid of the device and removing the
plates and/or the tubes.

EN

Operat

ing manual