IKA HB 4 basic User Manual
Page 13

The following error messages can be displayed (fig. 17) and have the fol-
lowing causes described:
No.
Error
Remedy
Er 3 Internal device temperature above 76°C Allow the device to cool down.
Mains switch OFF/ON
Er 4 Motor speed deviates ±300 rpm
Reduce load.
from the speed. Set speed greater Mains switch OFF/ON
than 350 rpm
Er 4 Stirring function turned on
Mains switch OFF/ON
Actual speed = 0 rpm
Increase the speed or turn the
Indication after approx 30 sec
stirring function
Er 20 Safety relay does not close
Safety circuit is set too low,
check mains voltage (mains
voltage must be larger than
the nominal voltage - 10%).
Mains switch OFF/ON
Er 21 Safety relay does not open
RESET key/service
Er 22 Disturbance before the last time the RESET button
device was switched OFF or safety
temperature is lower then the tem-
perature of the medium
Er 23 Adjustable safety circuit faulty
Mains switch OFF/ON / service
Er 24 Safety temperature set exceeded
Set safety temperature higher.
Mains switch OFF/ON
Er 26 System dry, no medium in the
Allow the device to cool down,
bath container
then refill with medium.
Mains switch OFF/ON
Er 27 Calibration error
Mains switch OFF/ON. When
switching on, a calibration is
automatically carried out.
Er 28 Sensor rupture
Mains switch OFF/ON
Er 29 Short circuit in the safety sensor or
Service
controller and safety sensor mixed up
Er 30 Short circuit in the controller sensor
Service
The safety circuit must be checked over at least once a year by the user.
To do this, fill the temperature contol bath with 1l of water and in opera-
ting mode „A“ set the safety temperature to 100°C. Then set the target
temperature to 80°C and start the heating function by pressing the Temp
ON/OFF key. After the target temperature has been reached, switch the
moderating bath off and back on again with the power switch and set the
safety temperature to 70°C. The temperature of the medium will then be
10°C above the safety temperature, and the safety circuit will be activa-
ted. Error message Er 22 or Er 24 must be displayed.
The aparate is maintenace-free. It is subject only to the natural wear
and tear of components and their statistical failure rate.
When ordering spare parts, please give the manufacturing num-
ber shown on the type plate, the machine type and the name of
the spare part.
Please send in equipment for repair only after it has been cleaned and
is free from any materials which may constitute a health hazard. Use
only cleansing agents which have been approved by IKA to clean IKA
devices. To remove use:
Dyes
isopropyl alcohol
Construction materials
water containing tenside / isopropyl alcohol
Cosmetics
water containing tenside / isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from IKA.
Wear the proper protective gloves during cleaning of the devices.
Electrical devices may not be placed in the cleansing agent for the pur-
pose of cleaning.
Before using another than the recommended method for cleaning or
decontamination, the user must ascertain with the manufacturer that this
method does not destroy the instrument.
13
HB4 0596
Error messages
Maitenance and cleaning
BA_HB4 30.06.2003 10:07 Uhr Seite 13