Technical data
Nominal voltage
VAC
230 ± 10% / 115 ± 10%
Frequency
Hz
50 / 60
Max. input power
W
300
Working temperature range
°C
-20 ... RT
Operating temperature range (with external heating) °C
-20 ... + 40
Temperature stability – internal temperature
control (according to DIN 12876)
K
± 0.1
Temperature control
PID
Temperature measurement, absolute accuracy
K
± 0.5
Temperature setting
Key on the front of the device
Temperature setting resolution
K
0.1
Temperature display
LED
Temperature display resolution
K
0.1
Classification according to DIN 12876-1
Class III (FL) suitable for flammable and non-flammable fluids
Cooling capacity according to DIN 12876
+ 20 °C
+ 10 °C
0 °C
- 10 °C
- 20 °C
W
400
370
320
240
130
Pump speed (adjustable)
rpm
1000 ... 3200
Max. pump pressure/suction
bar
0.3/0.2
Max. flow rate (at 0 bar)
l/min
15
Bath volume
I
1.4 – 4.0
Fluid maximum viscosity
mm
2
/s 50
Sub-level protection
Yes
Interface
USB, RS 232
Permitted on-time
%
100
Protection class according to EN 60 529
IP 21
Protection class
I
Excess voltage category
II
Contamination level
2
Permitted ambient temperature
°C
+ 5 ... + 32
Permitted ambient humidity
%
80
Dimension (W x D x H)
mm
220 x 525 x 475
Weight
kg
28
Operation at a terrestrial altitude
m
max. 2000