Maintenance and cleaning, Accessories – IKA T 65 digital ULTRA-TURRAX User Manual
Page 21

21
The device operates maintenance-free.
As necessary, but in any case not less than every six months, the
following points must be checked:
- EMERGENCY STOP switch operation
- Safety limit switch operation (with the stand in the lower position)
- Electrical consumers, connections and cables for damage.
Spare parts order
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part, see www.ika.com,
spare parts diagram and spare parts list
Repair
Please send instrument in for repair only after it has been
cleaned and is free from any materials which may consti-
tute a health hazard.
For this you should request the “Decontamination Clearance
Certificate” from IKA
®
, or use the download printout of it from
the IKA
®
website www.ika.com.
Return the instrument in its original packaging. Storage pa-
ckaging is not sufficient. Also, please use suitable shipping pa-
ckage materials.
Maintenance and cleaning
Only clean IKA
®
appliances using these IKA
®
approved cleaning
agents:
Dirt
Cleaning agent
Dyes Isopropanol
Building materials
Water containing
detergent,
Isopropanol
Cosmetics
Water containing
detergent,
Isopropanol
Food
Water containing
detergent,
Fuels
Water containing
detergent,
Other materials
Please consult IKA
®
- Wear protective gloves when cleaning the devices.
- Do not place electrical appliances into the cleaning agents for
cleaning purposes.
- Do not allow moisture to get into the appliance when cleaning.
- Please consult IKA
®
before using any cleaning or decontamina-
tion methods, other than those recommended here.
T 653
Telescopic stand
SI 400
Safety limit switch
SI 474
Support holder
Accessories
S 65 KG-HH-G 65 G
Dispersion tool coarse
S 65 KG-HH-G 65 M
Dispersion tool medium
S 65 KG-HH-G 65 F
Dispersion tool fine