beautypg.com

Thermocouple wire harness – J.P. Instruments EDM 730 Instrument Installation Manual User Manual

Page 10

background image

FAA APPROVED INSTALLATION MANUAL FOR THE EGT-701 Report 103


INSTALLING THE EGT-701 SCANNER®
1/20/09

Rev-E Page 10 of 30













Before you begin, note that the most common cause of probe problems is poor
terminal crimps.

When cutting each pair of leads in the wiring harness to the proper length to connect to the probes, leave enough
slack in the harness so that probes from adjacent cylinders may be interchanged if necessary for trouble-shooting and
servicing. Thermocouple wire length is not critical and should be trimmed to any length as required for a clean
installation.
The Temperature probes must be wired with the correct polarity. Each wire is marked with the cylinder number. The
EGT and CHT probes connect to the temperature indicator with yellow jacket Teflon Chromel Alumel wire supplied.

Strip the wires as shown below, observing color codes.

Important: fold back the wire double before

crimping terminals.

1/4"

1 1/2"

Fold back wire
double before
crimping terminals

2 1/4"

Thermocouple wire harness

red

yellow

Terminate each wire with a crimp-on ring terminal, provided. The ring terminals may be crimped with a “service-type”
tool, however AMP part number 48518 crimp tool is recommended. Verify the quality of each crimp with a sharp tug on
the wire. The terminal should be impossible to pull off when crimped correctly.

shrink tubing

shrink tubing

ring terminal

Place a ¼ x 4 inch sleeve over each pair of wires in the harness. Connect the harness ring terminal to the probe ring
terminal using the supplied number 4 screws and nuts, placing the star washer between the ring terminals, not against
the nut.

Important: place star waster between two ring
terminals and tighten nut and bolt as
necessary

to instrument

to probe

Slide the sleeve over the joint and secure with three tie-wraps.

1/4 x 4" sleeve

tie-wrap 3 places


Cut red lead if
transducer is
also connected
to another fuel
flow instrument

IN

OUT

WHT
BLK
RED

WHT
BLK
RED

fuel flow option

connector

1

1

9

8

15

FF
COM
EGT

+12/24 avionic bus

GPS

RS-232 input to EDM-700

GPS output

GPS input

(some KLN series only)

RS-232

output from

EDM-700

8
7
6
5
4
3
2
1

Fuel Flow Transducer
FloScan 201 -A, -B, -C, or 231

Alarm lamp or enunciator

(customer supplied option)

11

9

panel

mounted

switch

(alarm wire not supplied)

EGT

ALL

FF