( L ) EXTERNAL PROGRAMMING NOT OK.
EXTERNAL PROGRAMMING
NOT OK
Is Pot connected across INT
Ensure 5K Pot connected
& COM OK ?
between INT and COM
Terminal.
Is there proper Continuity between
PROG Terminal and O/P +Ve ?
Ensure proper Connectivity.
1.
Check whether 5K Pot is connected between INT and COM Terminal.
2.
Check whether there is proper Continuity between PROG Terminal and
Output Positive.
(M ) REMOTE SENSE NOT OK.
1.
Check whether there are Oscillations at Output, when Remote Sense Switch
is pressed. If Oscillations are present reduce Resistor R6 and R9 on Front
Panel PCB, until Oscillations die out.
N
N
Y
25
See also other documents in the category B&K Precision Power suppliers:
- XLN Series - Datasheet (5 pages)
- XLN3640 - Manual (207 pages)
- 9115-AT - Manual (74 pages)
- 9115 - Programming Manual (23 pages)
- 9115 Series - Datasheet (5 pages)
- 1760A - Manual (49 pages)
- 1760A - Datasheet (1 page)
- 1744A - Manual (28 pages)
- 1740B - Datasheet (1 page)
- 9120A - Manual (50 pages)
- 1672 - Manual (23 pages)
- 1692 - Manual (2 pages)
- 1685B - Manual (42 pages)
- 9110 - Manual (12 pages)
- 1665 - Manual (12 pages)
- 1710A - Manual (56 pages)
- 1626A-1627A - Manual (17 pages)
- 1686A - Manual (20 pages)
- 1671A - Manual (22 pages)
- 1670A - Manual (22 pages)
- 1550 - Manual (8 pages)
- 1680 - Manual (6 pages)
- 1514 - Manual (1 page)
- 1513 - Manual (1 page)
- 1796 - Manual (43 pages)
- 1795 - Manual (42 pages)
- 9170 - Manual (187 pages)
- 9130 - Manual (48 pages)
- 1747 - Manual (37 pages)
- 9120A - Datasheet (2 pages)
- 1672 - Datasheet (3 pages)
- 1747 - Datasheet (3 pages)
- 1737 - Datasheet (3 pages)
- 9130 - Datasheet (1 page)
- 9170 Series - Datasheet (6 pages)
- 1790 - Datasheet (1 page)
- 9150-9153 - Datasheet (2 pages)
- 1680 - Datasheet (1 page)
- 1604A - Datasheet (1 page)
- 1550 - Datasheet (1 page)
- 1670A - Datasheet (1 page)
- 1686A - Datasheet (1 page)
- 1623A - Datasheet (1 page)
- 1665 - Datasheet (1 page)