Page 15

background image

5.3.6 If Fuse blows, connect the unit through a variable AC Source with a current meter

and monitor the current at no load. If the current is very high (more than 2 ampere
or so ) load, check for Bridge Rectifier short or shorted Diode, or Secondary
Winding short, etc

5.3.7 If output voltage or current DPM's are not reading, check for loose connections

especially in crimping, soldering, of Connectors.

WARNING

THE FOLLOWING SERVICING INSTRUCTIONS ARE FOR USE BY

QUALIFIED PERSONNEL ONLY.TO AVOID PERSONAL INJURY, DO NOT

PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN

OPERATING INSTRUCTIONS UNLESS YOU ARE QUALIFIED TO DO SO.

-

REFER TO OPERATORS SAFETY SUMMARY AND SERVICE SAFETY SUMMARY PRIOR TO PER FOR MING A NY SERV ICE.