Liebherr WKt 4552 GrandCru User Manual
Page 30

30
LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller
9,0 °C
8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C
1,5 °C
1,0 °C
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
LCv 4010 – Evaluation of refrigerator compartment
H0 = Gradient
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
Gradient H0- :
6.0 K
(+2.6 K / – 3.4 K from
set temperature)
Max. fluctuation:
3.4 K
Gradient H0- :
3.4 K
(+1.8 K / –1.6 K from
set temperature)
Max. fluctuation:
4.0 K
Set temperature
Average probe temperature
LCv 4010 – Evaluation of freezer compartment
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
– 16,5 °C
– 17,0 °C
– 17,5 °C
– 18,0 °C
– 18,5 °C
– 19,0 °C
– 19,5 °C
– 20,0 °C
– 20,5 °C
– 21,0 °C
– 21,5 °C
– 22,0 °C
– 22,5 °C
– 23,0 °C
– 23,5 °C
– 24,0 °C
H0 = Gradient
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
Set temperature
Average probe temperature
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
- G 5216 FT 2900 WTpes 5972 Vinidor GGU 1500 Premium GGPv 1470 ProfiLine GGv 5060 BCDv 1003 WKt 4551 GrandCru WKb 4212 Vinothek WKr 1811 Vinothek Fv 3613 MKv 3910 ZKes 453 Humidor GGv 5860 FT 2902 FDv 4613 GGPv 1490 ProfiPremiumline WKes 4552 GrandCru WKes 653 Grand Cru GGv 5810 WTes 5972 Vinidor GG 4060 WTr 4211 Vinothek GGU 1550 Premium GG 5260 FKDv 3713 Premium BCDv 4313 WTb 4212 Vinothek GGv 5010 WKt 6451 GrandCru GG 4010 WKb 3212 Vinothek WTes 5872 Vinidor MKUv 1610 GGPv 6570 ProfiLine WKt 5552 GrandCru WKr 4211 Vinothek WKb 1812 Vinothek GG 5210 GGPv 6590 ProfiPremiumline WKr 3211 Vinothek WKt 5551 GrandCru