Temperature consistency and distribution
according to NF X 15-140
The Liebherr laboratory cabinets are all tested in-house to the NF X 15-140 test procedure. Liebherr’s modern R & D facility has climate controlled test rooms
which conform to EN 23953.
Description of the NF X 15-140 test procedure
1
2
3
4
6
7
8
9
5
centred
1/10 B
1/10 T
1/10 H
LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller
The general test conditions are as follows:
■
Ambient temperature + 25 °C
■
60 % relative humidity
■
Refrigerator set temperature + 5 °C
■
Freezer set temperature – 20 °C
■
Low temperature freezer set temperature – 45 °C
Measurements are made using 9 x PT 100 air temperature probes positioned in accordance with NF X 15-140.
“Unless specifically stipulated, the temperature probes are located at equal distances from the walls of 1/10 of each of the dimensions of the internal volume
(w/h/d). The location of the probes used for characterisation forms a supposedly representative sample of the working space.”
The characterisation is performed using nine probes; one located at each corner, one in the centre of the working space.
Location of the probes used for characterisation
- G 5216 FT 2900 WTpes 5972 Vinidor GGU 1500 Premium GGPv 1470 ProfiLine GGv 5060 BCDv 1003 WKt 4551 GrandCru WKb 4212 Vinothek WKr 1811 Vinothek Fv 3613 MKv 3910 ZKes 453 Humidor GGv 5860 FT 2902 FDv 4613 GGPv 1490 ProfiPremiumline WKes 4552 GrandCru WKes 653 Grand Cru GGv 5810 WTes 5972 Vinidor GG 4060 WTr 4211 Vinothek GGU 1550 Premium GG 5260 FKDv 3713 Premium BCDv 4313 WTb 4212 Vinothek GGv 5010 WKt 6451 GrandCru GG 4010 WKb 3212 Vinothek WTes 5872 Vinidor MKUv 1610 GGPv 6570 ProfiLine WKt 5552 GrandCru WKr 4211 Vinothek WKb 1812 Vinothek GG 5210 GGPv 6590 ProfiPremiumline WKr 3211 Vinothek WKt 5551 GrandCru