Gradient: “Distribution of temperatures within the chamber“
H0
= max (0
-
mj
+U
mj
) – min (
0
-
mj
-U
mj
)
max = (mean value warmest probe + expanded measurement uncertainty)
min = (mean value coldest probe – expanded measurement uncertainty)
Cabinets are tested empty over a period of 24 hrs including defrost cycles.
Test criteria for characterising the temperature consistency and distribution
1)
Gradient or homogeneity: “The gradient H0 is the maximum difference, obtained in an achieved environment, between the mean values of measurements 0
mj
increased by their expanded uncertainty U
mj
, within the working space during the measurement time.”
Example: Gradient (Homogeneity)
The gradient indicates the temperature distribution within the cabinets. Especially in the laboratory sector optimum temperature distribution is crucial to guarantee
the safe storage of samples or other products at any place within the unit. The lower the gradient the better is the temperature distribution within the unit.
Gr
adient H0
LKPv and LGPv laboratory appliances with Profi-Controller
Maximum value
Minimum value
Measurement period
- G 5216 FT 2900 WTpes 5972 Vinidor GGU 1500 Premium GGPv 1470 ProfiLine GGv 5060 BCDv 1003 WKt 4551 GrandCru WKb 4212 Vinothek WKr 1811 Vinothek Fv 3613 MKv 3910 ZKes 453 Humidor GGv 5860 FT 2902 FDv 4613 GGPv 1490 ProfiPremiumline WKes 4552 GrandCru WKes 653 Grand Cru GGv 5810 WTes 5972 Vinidor GG 4060 WTr 4211 Vinothek GGU 1550 Premium GG 5260 FKDv 3713 Premium BCDv 4313 WTb 4212 Vinothek GGv 5010 WKt 6451 GrandCru GG 4010 WKb 3212 Vinothek WTes 5872 Vinidor MKUv 1610 GGPv 6570 ProfiLine WKt 5552 GrandCru WKr 4211 Vinothek WKb 1812 Vinothek GG 5210 GGPv 6590 ProfiPremiumline WKr 3211 Vinothek WKt 5551 GrandCru