Temperature and alarm documentation – Liebherr WKt 4552 GrandCru User Manual
Page 19

Address 1
Address 2
Address 3
RS 485
+ - s
- + s
Optimum:
Shielded Twisted Pair
- + s
- + s
Terminator
120 Ohm
Terminator
120 Ohm
BUS line
appliance 1 appliance 2 appliance 20
Liebherr
Temperature
Monitoring
Software for
Serial Interface &
Infrared key
Deutsch
English
Français
Italiano
Español
19
External temperature and alarm
monitoring connection
RS 485 interface and optional signal converter
All LK(U)v, LCv, LK(U)exv and LG(U)ex models have an integrated RS 485 interface. If an RS 485 monitoring
system is not already in place, a converter kit is available as an optional extra to convert the RS 485 signal
to an RS 232 signal and allow connection to a PC or laptop.
The RS 485 bus connection is made using standard shielded data cables, e.g. type LiYCY 2 x 0,14 mm². The
maximum possible length of the data lines depends on the quality of the components used. The maximum
range is 500m. A maximum of 20 appliances can be wired in series.
Dedicated software is provided in the accessory package allowing visualisation, analysis and export to other
formats for storage. The software reads the air-probe temperature at 2 minute intervals and records alarm
conditions. When an alarm condition is registered the software will warn on the PC on which the software
is running: if this PC is not supervised (e.g. at night or over weekends) then the alarm signal might not be
recognised. In order to avoid unrecognised alarms the software includes an email function, which sends an
email in case of an alarm condition to predefined email addresses. These emails can be easily checked with
a smartphone. Therefore the user will be warned without having direct access to a computer.
Networking
Both laboratory appliances with Profi-Controller as well as laboratory appliances with Comfort- Controller
can be networked using the same Liebherr accessory package “RS 485 interface converter including
temperature monitoring software”.
LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller
Temperature and
alarm documentation
Liebherr Mediline refrigerators and freezers with Comfort-Controller help to protect your inventory. These cabinets offer a variety of features
and options designed both for external monitoring of the internal temperatures, the product temperatures and alarm conditions to greatly
reduce the risk of loss of contents.
Example: RS 485 network
- G 5216 FT 2900 WTpes 5972 Vinidor GGU 1500 Premium GGPv 1470 ProfiLine GGv 5060 BCDv 1003 WKt 4551 GrandCru WKb 4212 Vinothek WKr 1811 Vinothek Fv 3613 MKv 3910 ZKes 453 Humidor GGv 5860 FT 2902 FDv 4613 GGPv 1490 ProfiPremiumline WKes 4552 GrandCru WKes 653 Grand Cru GGv 5810 WTes 5972 Vinidor GG 4060 WTr 4211 Vinothek GGU 1550 Premium GG 5260 FKDv 3713 Premium BCDv 4313 WTb 4212 Vinothek GGv 5010 WKt 6451 GrandCru GG 4010 WKb 3212 Vinothek WTes 5872 Vinidor MKUv 1610 GGPv 6570 ProfiLine WKt 5552 GrandCru WKr 4211 Vinothek WKb 1812 Vinothek GG 5210 GGPv 6590 ProfiPremiumline WKr 3211 Vinothek WKt 5551 GrandCru