6 search auto zero – Bacharach H25-IR User Manual
Page 23

H25-IR
Preparing for Operation
Instruction 3015-4342
17
2.6.6 Search Auto Zero
The SEARCH AUTO ZERO function allows the operator to select one of the following conditions when the auto-
zero function is disabled while in the Search mode:
NOTE: This function works in conjunction with the setpoint value as entered in Section “2.6.5 Setpoint.”
ALWAYS ON – The background gas level is always zeroed. If the probe is held over a gas leak, the instrument
will automatically zero-out the reading.
ALWAYS OFF – The background gas level is never zeroed. If the probe is held over a gas leak, the instrument
will not zero-out the reading, similar in operation to the Measure mode.
BELOW SETPOINT – The auto-zero function is ON when the leak-rate is below the setpoint value. If the leak
rate goes above the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay on the
leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <2 Oz/yr. If
the leak rate is
2 Oz/yr, then the auto-zero function turns OFF.
BELOW 90% OF SP – The auto-zero function is ON when the leak-rate is below 90% of the setpoint value. If
the leak rate goes above 90% of the setpoint value, then the auto-zero function turns OFF, allowing the probe
to stay on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <1.8 Oz/yr.
If the leak rate is
1.8 Oz/yr, then the auto-zero function turns OFF.
BELOW 75% OF SP – The auto-zero function is ON when the leak-rate is below 75% of the setpoint value. If
the leak rate goes above 75% of the setpoint value, then the auto-zero function turns OFF, allowing the probe
to stay on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <1.5 Oz/yr.
If the leak rate is
1.5 Oz/yr, then the auto-zero function turns OFF.
BELOW 50% OF SP – The auto-zero function is ON when the leak-rate is below 50% of the setpoint value. If
the leak rate goes above 50% of the setpoint value, then the auto-zero function turns OFF, allowing the probe
to stay on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <1 Oz/yr. If
the leak rate is
1 Oz/yr, then the auto-zero function turns OFF.
SEARCH AUTO ZERO
SETUP NO.1
<
ALWAYS ON
ALWAYS
OFF
BELOW
SETPOINT
BELOW 90% OF SP
BELOW 75% OF SP
BELOW 50% OF SP