Cleaning the cuff and recorder

Before cleaning the recorder, remove the battery cover and turn the power

switch off. Remove the batteries.

The recorder is not water resistant, do not allow liquids to splash on or get into

the case while cleaning.

After each use, wipe the case of the recorder with a clean lint free cloth,

moistened with water and a mild detergent.

Do not use antiseptic solutions, Alcohol, etc., to clean the recorder, hose or cuff.

Clean the cuff cloth and cuff cover by washing in water with a mild detergent.

Do not scrub or wring them by hand. If the cuff cloth and cuff cover become

contaminated, replace them with new covers.

Periodical inspection

This blood pressure recorder is a precision instrument. Please inspect the

functions (every year) periodically. Contact your nearest A&D office for this

inspection.

Problem solving

Caution

Do not open the case of the recorder because it uses delicate electrical

components and an intricate air unit that could be damaged.

If you can not locate and fix the problem, request service from your supplier,

or from the A&D service group.

A&D service group will support authorized suppliers with technical informa-

tion, spare parts and units.

Maintenance

Cleaning the Cuff and Recorder

Problem

No display at turn-

ing on.

Data lost while re-

placing batteries.

No pressure.

Cause

Battery power has been

consumed.

Unable to charge the in-

ternal sub battery.

Air leakage at the con-

nector, hose or cuff.

Treatment

Replace with new batteries

Set monitor for clock display

mode for approx. 24 hours. Do

not take BP measurement. The

sub-battery battery charges dur-

ing clock display mode.

Confirm the cuff and air hose

are not damaged and are con-

nected correctly.