Features, Calibration, Parts description – MyWeigh iBALANCE 201 User Manual
Page 3: When to calibrate - calibration is rarely required, How to calibrate
CALIBRATION
When to calibrate - calibration is RARELY required.
Calibration may be required when the scale is first set up for use, or if the sca le is
moved to a different altitude or gravitati on. This is necessary because the weight
of a mass in one location is not necessarily the same in another location. Also, with
time and use, mechanical deviations can occur.
How to calibrate
**you must have an accurate 100 gr am weight or combination of weights in
order to calibrate**
1. The scale must be power ed OFF. Press and hold [ZERO] first, then P ress
[ON/OFF] while keeping the [ZERO] button depressed, then re lease both key s.
The display will show the A/D value (a series of random numbers).
2. Wait for the stable indicator to be displayed , Press [UNIT] the display will show
"00SAVE", then after 2 seconds the display will return to the random A/D value.
3. Place 100 gram s of weight(s) on p latform. W ait for the stable indic ator to
display, th en press [UNIT], th e display will sho w "00SAVE". Calibration i s
complete, remove the weight(s).
4. Turn the s cale OFF (Pres s the ON/OFF key) , then turn it back O N and che ck
some weight readings. If calibration is still incorrect, repeat calibration but try it
more slowly.
If calibration is repeatedly incorrect then: 1) Re place the batteries, 2) Try a more
stable surface or vi bration and i nterference free lo cation, 3) Be su re there is no
static electricity on the scale (use an anti-static spray to remove static electricity)
FEATURES
z
Auto shut off (optional)
●
Auto calibration
z
Auto zero tracking
●
Auto backlight (optional)
z
Low battery indication
●
g /oz/mg/ozt/dwt/ct/gn/pcs conversion
z
Large LCD (6digits 15 mm high)
●
with counting function
z
Large square weighing tray
z
Net weight/stability indication
z
0.005 gram division – professional accuracy
5
PARTS DESCRIPTION
RIGHT SIDE
AC adaptor
Socket
TO P
Weighing platform
The arch face of platform is front.
Correct the direction of the pan.
Display
ON/OFF key
Turns the scale
power on/off.
PCS key
UNIT key count ing mode
Changes weighing units, g ,oz,lb, ozt, dwt, ct,gn or pcs(counting mode).
ZERO key (for TARE)
ets display to zero or
Subtracts weight of a container.
BO TTOM
Battery compartment
Battery compartment cover
Batteries
Use 6 AA batteries
DISPLAY
STABLE indicator Auto backlight function indicator
Indicates when the reading is stable.
Auto shut off function indicator
ZERO indicator
Indicates when the reading is stable
Weighing unit
Low battery indicator
2