7 Maintenance and cleaning

22

WIKA operating instructions, model A-AI-2

You have to set:

offset =

0.08

(= zero point-deviation)

scale =

0.30

(= deviation/actual-slope=0.06/9.94=0.0030= 0.30%)

6.5 Min-/max-value

storage

The device features a minimum/maximum-value storage. In this storage the highest and lowest
performance data is saved. The storage is non-permanent so the saved values get lost in case of
voltage drop.

Calling of the minimum-value

press button 3 shortly

the device will display “Lo
briefly, after that the min-value is
displayed for about 2 sec

Calling of the maximum-value

press button 2 shortly

the device will display “Hi
briefly, after that the max-value
is displayed for about 2 sec.

Erasing of the min/max values

press button 2 and 3 for 2 sec.

The device will display “CLr
briefly, after that the min/max-
values are set to the cur-rent
displayed value.

7 Maintenance

and

cleaning

7.1 Maintenance

This instrument is maintenance-free.

Repairs must only be carried out by the manufacturer.

7.2 Cleaning



CAUTION!

Before cleaning, correctly disconnect the instrument from the mains.
Clean the instrument with a moist cloth.
Electrical connections must not come into contact with moisture.


For information on returning the instrument see chapter "9.2 Return".