Failure

Possible cause

Procedure

Output signal unchanged after change

in pressure

Mechanical overload through over-

pressure

Replace instrument; if failure reoccurs,

consult the manufacturer *)

Wrong supply voltage or current spike Replace instrument

No/False output signal

Incorrectly wired (e.g. Connected as

2-wire instead of 3-wire system)

Follow pin assignment (see Instrument

Label / Operating Instructions)

No output signal

No/incorrect voltage supply or current

spike

Adjust the voltage supply to correspond

with the Operating Instructions *)

Cable break

Check connections and cable

Signal span too small

Power supply too high/too low

Correct the power supply in line with

the Operating Instructions

Signal span too small

Mechanical overload through over-

pressure

Replace instrument; if failure reoccurs,

consult the manufacturer *)

Signal span drops off

Moisture present (e.g. at the cable tail) Install the cable correctly

Signal span erratic

Electromagnetic interference source in

the vicinity, e.g. inverter drive

Shield the device; shield the cables;

remove the interference source.

Instrument not grounded

Ground instrument

Strong fluctuations in the power supply Stabilise the power supply; smooth it

(i.e.; remove interferences)

Violent fluctuations in the process

media pressure

Damping; consult with manufacturer

9. Trouble shooting

GB

In case of unjustified reclamation we charge the reclamation handling expenses.
*) Make sure that after the assembly the unit is working properly. In case the error continues to exist send in the instrument for

reparation (or replace the unit).

Please verify in advance if pressure is being applied (valves / ball valve etc. open) and if the
right voltage supply and the right type of wiring (2-wire / 3-wire) has been chosen?

If the problem persists, contact our sales department.
USA, Canada
If the problem continues, contact WIKA or an authorized agent for assistance. If the pres-
sure transmitter must be returned obtain an RMA (return material authorization) number and
shipping instructions from the place of purchase. Be sure to include detailed information about
the problem. Pressure transmitters received by WIKA without a valid RMA number will not be
accepted.

Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.wika.de / www.wika.com).

9. Trouble shooting / 10. Storage, disposal

GB

10. Storage, disposal

When storing or disposing of the pressure transmitter, take precautions with
regard to remaining media in removed pressure transmitters. We recom-
mend cleaning the transmitter properly and carefully. Remaining media in
the pressure port may be hazardous or toxic!

Dispose of instrument components and packaging materials in accordance with the
respective waste treatment and disposal regulations of the region or country to which
the instrument is supplied.

Disposal

WIKA reserves the right to alter these technical specifications.

!

Warning