Myron L 6PFCE and 4P User Manual
Page 25
![background image](/manuals/460915/25/background.png)
Figure 7
BUFFER
pH
CAL
Figure 8
pH
BUFFER
CAL
NOTE: If the “Acd” and “bAS” indicators are blinking, the unit is indicating
an error and needs either an acid or base solution present in the sensor
well.
3.
Rinse sensor well 3 times with acid or base buffer solution.
4.
Refill sensor well again with same buffer solution.
5.
Press
or
until display agrees with buffer value.
6.
Press
to accept 2nd point of calibration. Now the
display indicates the next type of buffer to be used.
Single point Gain Calibration is complete. You may continue for the 3rd
point of Calibration (2nd Gain) or exit by pressing any measurement key.
Exiting causes the value accepted for the buffer to be used for both acid
and base measurements.
To continue with 3rd point calibration, use basic buffer if acidic buffer
was used in the 2nd point, or vice-versa. Again, match the display to the
known buffer value as in step 2 and continue with the following steps:
7.
Repeat steps 3 through 6 using opposite buffer solution.
8.
Press
to accept 3rd point of calibration, which
completes the Calibration procedure. Fill sensor well with
Myron L Storage Solution and replace protective cap.
6. ORP/FC
E
Calibration (6PFC
E
)
ORP electrodes rarely give false readings without problems in the
reference electrode. For this reason, and because calibration solutions
for ORP are highly reactive and potentially hazardous, your Ultrameter II
has an electronic ORP calibration. This causes the zero point on the
reference electrode to be set whenever pH 7 calibration is done.
7. Temperature Calibration
Temperature calibration is not necessary in the Ultrameter II.
21